Technical Specifications

DETECTOR PHYSICAL SPECIFICATIONS

Number of Detectors

2, Fixed 90° Angle

UFOV ( W x H )

14.8" x 9.5" ( 37.6 x 24.1 cm)

Crystal Thickness

3/8" (9.5mm)

Number of PMTs / Detector

35

Acquisition

1 ADC per PMT

Energy Range

50 - 200 keV

Collimators

2, LEHR

DETECTOR PERFORMANCE SPECIFICATIONS

Intrinsic Spatial Resolution

≤ 3.5 mm

Intrinsic Spatial Linearity

≤ 0.5 mm differential

Intrinsic Energy Resolution

≤ 9.7% FHWM

System Resolution FWHM @ 10 cm - without scatter

8.6 mm - LEHR

System Resolution FWHM @ 10 cm - with scatter

10.0 mm - LEHR

System Sensitivity

260 cpm/uCi - LEHR

GANTRY SPECIFICATIONS

Height

50.3" (127.8cm)

Width

32.5" (82.6cm)

Length

55.6" (141.2cm)

Weight

1050 lbs. (476 kg)

90 degree turn radius

48”

Maximum supported patient weight

500 lbs (226.8 kg)

Chair seat to floor distance

24” @ lowest chair position

Chair vertical motion range

10.0" (25.4cm)

Detector rotational speed

0 to 1.0 RPM

Min/max circular orbit

4.0-12.0" radius (10.2 - 30.5cm)

Patient contouring

Automatic, 2 learning points

Min/max non-circular orbit

3.5-13.8" radius (8.9-35.0 cm)

ENVIRONMENTAL SPECIFICATIONS

Power requirements

120VAC, 60Hz, 10Amps max.

Operating temperture range

60-86°F (16-30°C)

Humidity range (non-condensing)

30-80%

Max. temperature gradient

8°F/hr (4.4°C/hr)

Min. recommend floor space

5’ x 7’ (152.4 x 213.4cm)

Floor leveling

Flat surface, level, even floor

INTEGRATED ACQUISITION/PROCESSING CONSOLE

Operating system

Windows XP Professional

Multi-functional color monitor

17" LCD

Persistence mode

Included

ECG trigger gate

Included (3-lead)

Tomographic acquisitions

SPECT and gated SPECT

Planar acquisitions

Static, Dynamic, and Gated

Processing interface

DICOM 3.0

Hardcopy

Network Postscript, DICOM

Multitasking

Simultaneous Acquisition & Processing