Technical Specifications
Imaging Performance Specifications
Energy Range and Resolution |
Energy Range: 90-160 Kev
Intrinsic Spatial Resolution: 4 mm
|
Intrinsic Spatial Linearity |
Differential: 0.3 mm CFOV, 0.4 mm UFOV |
Intrinsic Flood Uniformity |
Integral: 4% CFOV, 5% UFOV Differential: 2.5% CFOV, 3.5% UFOV |
System Spatial Resolution Without Scatter |
10 mm FWHM, 20 mm FWTM @ 100 mm |
Intrinsic Count Rate Performance in Air |
20% Count Loss rate of ≥ 250 Kcps (observed) / 200 Kcps (input) and a maximum count rate > 500 Kcps. |
Computer Specifications
|
15" LCD monitor for P-Scope and EKG display |
|
20" LCD monitor at processing station |
|
High performance Pentium 4 CPU running a stable Linux operating system |
|
DVD multi session archival storage |
|
512 MB DDR SDRAM |
Dual Monitor Graphics Adapter |
|
CD-RW Drive |
|
40 GB Hard Drive |
|
Integrated Network Adapter |
|
Multimedia Keyboard |
|
Optical Mouse |
System Features
|
Wide area network connection enables remote system diagnostics, software updates, and automated backup of system files |
|
AutoStart feature begins acquisition when the bolus is detected |
|
Complete FPRNA processing and review suite including motion correction |
|
Remote sharing software allows customer support to assist with and observe the customer processing in real time |
|
WebReview enables the physician to view studies using a standard internet browser |
Camera Specifications
|
Total System Weight: 400 lbs |
|
Operating Temperature: 68-80F, Humidity: 20-80% non-condensing |
|
System Footprint: 20.5" x 51.5" |
|
Power Requirements: 110/120 Volt AC 15 amp dedicated circuit |
|
Installs on either side of treadmill |
|
Detector UFOV: 8.5" x 8.5" |
Collimators Available: Parallel or Slant-hole |
|
Detector Vertical Motion Range at CFOV: 28" to 67" |
|
Medical Grade UPS power failure protection |
|
System Demonstrated Uptime: > 99% |
|
Acquisition: 30 second list mode |